HBP-1320 Professional Blood Pressure Monitor

Technical Specifications
Technical Specifications
Product CategoryElectronic Sphygmomanometers
Product descriptionAutomatic Upper Arm Blood Pressure Monitor
Model (code)HBP-1320 (HBP-1320-E)
Measurement ParameterNIBP, PR
DimensionMain unit: 123 x 201 x 99 (mm) 4.84 x 7.91 x 3.90 (inch) (W x H x D)
AC adapter: 46 x 66 x 66.2 (mm) 1.81 x 2.60 x 2.61 (inch) (W x H x D)
Battery: 54 x 43.5 x 15.4 (mm) 2.13 x 1.71 x 0.61 (inch) (W x H x D)
WeightMain unit: Approx. 0.52 kg (not including accessories and options)
AC adapter: Approx. 114 g
Battery: Approx. 0.1 kg
Display7 segment LCD
Protection ClassClass II (AC Adapter)
Internal powered equipment (when operating with battery only)
Degree of ProtectionType BF (Applied part): Cuff
Operation ModeContinuous operation
MDR ClassificationClass II a

 

Power supply
AC adapterInput voltage range: AC 100 V to 240 V
Frequency: 50/60 Hz
Output voltage range: DC 6 V ±5%
Rated output current: 1.6 A
Rechargeable batteryType: 3.6 V, 1900 mAh
Number of operation cycles when fully charged: 300
Measurement conditions
• New battery fully charged
• Ambient temperature of 23°C (73.4°F)
• Using M-size cuff
• SYS 120 / DIA 80 / PR 60 (Inflation setting: AUTO)
• One 5-minute cycle consisting of “cuff measurement time + wait time”

 

Environmental Conditions
Operational temperature and humidityTemperature range: 5 to 40°C (41 to 104°F)
Humidity range: 15 to 85%RH (not condensed)
Atmospheric pressure: 700 to 1060hPa
Storage and transportationTemperature range: -20 to 60°C (-4 to 140°F)
Humidity range: 10 to 95%RH (not condensed)
Atmospheric pressure: 500 to 1060hPa

 

Non-Invasive Blood Pressure (NIBP)
Measurement technologyOscillometric
Measurement methodDynamic Linear Deflation method
Pressure display range0 to 300 mmHg
Pressure display accuracyWithin ±3 mmHg
NIBP measurement rangeSYS 60 to 250 mmHg
DIA 40 to 200 mmHg
Pulse rate 40 to 200 /min
NIBP accuracyMaximum mean error within ±5mmHg
Maximum standard deviation within 8mmHg
Pulse rate accuracyWithin ±5 % of reading